SustaPEEK MG was developed for the demanding requirements of the healthcare industry in particular medical devices. As a material of highest purity with exceptional resistance to sterilisation, it is e.g. ideal for surgical instruments. Thanks to its very high strength and rigidity, SustaPEEK MG ensures reliable performance in all applications. It retains its shape and property profile even under extreme conditions. With its excellent temperature stability of up to 250 °C in continuous use, it offers reliable performance.
We are a system supplier and partner from the idea to the OEM's end product - as a cooperative value contribution. We are able to support the healthcare industry at the highest level.
Armin Reuner - Industry Manager Healthcare - Mail: AReuner@roechling.com
Test Method | Unit | Guideline Value | |
---|---|---|---|
Density | ASTM D792 | g / cm3 | 1.32 |
Water Absorption 24 hours | ASTM D570 | % | 0.1 |
Water Absorption Saturation | ASTM D570 | % | 0.5 |
Test Method | Unit | Guideline Value | |
---|---|---|---|
Hardness | ASTM D2240 | Shore D | 85 |
Tensile Strength at yield 73°F | ASTM D638 | psi | 16000 |
Tensile Modulus | ASTM D638 | psi | 500000 |
Elongation at Break | ASTM D638 | % | 20 |
Flexural Strength | ASTM D790 | psi | 25000 |
Flexural Modulus | ASTM D790 | psi | 600000 |
Rockwell Hardness | ASTM D785 | 105 | |
Rockwell Hardness | ASTM D785 | R | 126 |
Izod Impact, Notched | ASTM D256 | ft-lb/in | 1.2 |
Test Method | Unit | Guideline Value | |
---|---|---|---|
Thermal Conductivity | in/hr/ft2 /°F | 1.73 | |
Coefficient of Linear Thermal Expansion | ASTM D696 | in/in/°F x10-5 | 2.6 |
Melting Point | ASTM D789 | °F | 630 |
Continuous Service Temperature, Air | °F | 480 | |
Deflection Temperature at 1.8Mpa (264psi) | ASTM D648 | °F | 320 |
Deflection Temperature at 1.8Mpa (66psi) | ASTM D648 | °F | 360 |
Test Method | Unit | Guideline Value | |
---|---|---|---|
Surface resistivity | ASTM D257 | Ω/cm | >1013 |
Test Method | Unit | Guideline Value | |
---|---|---|---|
FDA | Yes | ||
NSF | No | ||
USDA | Yes |